Model: NL-TCBS-4
The problem
The boiling water resistance of vitreous enamel on water heater tanks and enamel powder cannot be settled by eye, and batch differences run wide. The standard method demands a controlled boiling soak with quantitative weighing normalised by contact area; cooking samples on a stove never produces reproducible data, and supplier audits and type testing both call for a standard rig.
Capability
The NL-TCBS-4 runs four stations in parallel, each with independent temperature control and independent timing, so four recipes or sample sets are compared in one batch. The test follows ISO 28706-2: four 10 x 10 mm enamel specimens are numbered, weighed and mounted in boiling water glass vessels; the borosilicate glass meets GB/T 12416.2 chemical stability with thermal shock difference above 120C. A snake shaped reflux condenser with 400 mm jacket and standard ground joints returns condensate at 6 to 10 ml every 3 minutes. Each station takes 450 ml of test solution with specimens up to 110 mm diameter; the power is trimmed the moment the first bubbles appear and the gentle boil is held for 24 hours at a stretch. The full run lasts 504 hours, 21 days, with fresh water each day, after which the specimens are reweighed and the weight loss is computed as A = B / 8 / C, reproducible and archivable. SV timers accept minutes and seconds 0-99 and hours 0-999 on a PLC touch screen. The test durations cover boiling citric acid 2.5 h, sulfuric acid 18 h, hydrochloric acid 7 days and distilled water 48 h. A 40x magnifier is included for surface inspection, and the rig draws only 500 W at 220 V.


Applicable standards
ISO 28706-2 Vitreous and porcelain enamels, determination of resistance to chemical corrosion, part 2 (current); GB/T 20289-2006 storage water heaters, boiling water requirements (current); GB/T 4706.1-2024 household and similar appliances safety part 1 (current); QB/T 2590-2021 vitreous enamel parts of storage water heaters (current); loss thresholds follow the standards and the customer acceptance agreement.
Key specifications
- Stations: 4, independent temperature control and timing
- Standards: ISO 28706-2 / GB/T 20289-2006 / QB/T 2590-2021
- Specimens: enamel plates 10 x 10 mm, 4 pieces; diameter up to 110 mm
- Solution: 450 ml per station
- Vessel: borosilicate glass (GB/T 12416.2, thermal shock above 120C)
- Condenser: snake coil reflux type, 400 mm jacket, ground joints
- Condensate: 6 to 10 ml per 3 minutes
- Heating: gentle boil, power trimmed at first bubbles
- Cycle: 504 h (21 days), water renewed daily
- Timer: minutes and seconds 0-99 / hours 0-999
- Weight loss: A = B / 8 / C
- Media: citric acid 2.5 h / sulfuric 18 h / hydrochloric 7 days / distilled 48 h
- Display: PLC touch screen
- Inspection: 40x magnifier included
Applications
Incoming boiling water testing of water heater enamel tanks and enamel powder; type testing of vitreous enamelled ware; enamel recipe comparison trials; enamel corrosion testing at third party laboratories
About the manufacturer
Guangdong Endurance Electronic Equipment Co., Ltd provides non-standard customization, drawing confirmation, factory acceptance, on-site installation and commissioning, operation training and long-term technical service; 12-month warranty and long-term spare parts supply. Tel: 0760-88500902 / 13416787649. Web: en.zsnaili.com.